Platinum
Element density: 21.09 g/cm3
Electron configuration: [Xe]4f145d96s1
Oxidation states: +2,4

Edges
Explore electron energy loss spectroscopy (EELS) reference spectra, featuring major and minor edges used for elemental identification and chemical state analysis in materials research.
Major edges | |||
|---|---|---|---|
L3 | 11564 |
|
|
M4 | 2202 |
| |
M5 | 2122 |
| |
Minor edges | |||
|---|---|---|---|
L1 | 13880 |
|
|
L2 | 13273 |
|
|
M1 | 3296 |
|
|
M2 | 3027 |
| |
M3 | 2645 |
| |
N1 | 722 |
| |
N2 | 609 |
| |
N3 | 519 |
| |
N4 | 331 |
| |
N5 | 313 |
| |
N6 | 74 |
| |
N7 | 71 |
| |
O1 | 102 |
| |
O2 | 65 | ||
O3 | 52 | ||
Considerations
Pt has a strong M4,5-edge that shows both a strong cross-section and typically low background due to its high energy. To efficiently measure this edge, a large collection angle must be used. Some TEMs do not support high energy loss at short camera lengths due to cut off apertures in the column. Sudden drops in the power law background at high energy typically indicate cut off in the column.
Enhance your EELS community
Share spectra with your fellow EELS users from around the world.
Related spectral data

Pt thin film (0 – 170 eV)
Material analyzed: Pt
Related elements: Platinum
Data contributor: Gatan
Beam energy (keV): 200
Nominal energy range (eV): 0 – 170
Collection semi-angle effective (mrad): 100
Notes: Serial EELS; 20 sweeps 20 ms dwell; all analog

Pt thin film (0 – 3870 eV)
Material analyzed: Pt
Related elements: Platinum
Data contributor: Gatan
Beam energy (keV): 200
Nominal energy range (eV): 0 – 3870
Collection semi-angle effective (mrad): 100
Notes: Serial EELS; zero-loss peak intensity = 3.88e9; total elastic int. / zero-loss int. = 0.21; gain change factor = 302; 75 sweeps 10 ms dwell; saturated zero-loss
Applications
Unlock the power of electron energy loss spectroscopy—explore cutting-edge applications, dive into expert-led webinars, and discover insights through detailed application notes that fuel scientific breakthroughs.