Lead

Element density: 11.34 g/cm3
Electron configuration: [Xe]4f145d106s26p2
Oxidation states: +2,4

Lead

Edges

Explore electron energy loss spectroscopy (EELS) reference spectra, featuring major and minor edges used for elemental identification and chemical state analysis in materials research. 
 

 

Major edges

L3

13035

 

 

M4

2586

 

Major_Delayed Minimum

M5

2484

 

Major_Delayed Minimum

O4

22

Minor_Sharp Peak
Major_Delayed Minimum

O5

19

Minor_Sharp Peak
Major_Delayed Minimum

 

Minor edges

L1

15861

 

 

L2

15200

 

 

M1

3851

 

 

M2

3554

 

Very Weak_Delayed Minimum

M3

3066

 

Minor_Delayed Minimum

N1

894

 

Very Weak_Abrupt Onset

N2

764

 

Very Weak_Delayed Minimum

N3

645

 

Very Weak_Delayed Minimum

N4

435

 

Minor_Very Delayed

N5

413

 

Minor_Very Delayed

N6

143

 

Very Weak_Very Delayed

N7

138

 

Very Weak_Very Delayed

O1

147

 

Very Weak_Abrupt Onset

O2

105

Minor_Broad Peak
Major_Delayed Minimum

O3

86

Major_Broad Peak
Major_Delayed Minimum

Considerations

The M4,5-edge that shows both a strong cross-section and typically low background due to its high energy. To efficiently measure this edge, a large collection angle must be used. Some TEMs do not support high energy loss at short camera lengths due to cut off apertures in the column. Sudden drops in the power law background at high energy typically indicate cut off in the column.

Enhance your EELS community

Share spectra with your fellow EELS users from around the world.

Related spectral data

Pb electropolished foil (0 – 180 eV)
   

Pb electropolished foil (0 – 180 eV)

Material analyzed: Pb

Related elements: Lead

Data contributor: Gatan

Beam energy (keV): 200

Nominal energy range (eV): 0 – 180

Collection semi-angle effective (mrad): 100

 

Notes: Serial EELS; slightly oxidized; 10 sweeps 16.67 ms dwell

 

Pb electropolished foil (0 – 3890 eV)
   

Pb electropolished foil (0 – 3890 eV)

Material analyzed: Pb

Related elements: Lead

Data contributor: Gatan

Beam energy (keV): 200

Nominal energy range (eV): 0 – 3890

Collection semi-angle effective (mrad): 100

 

Notes: Serial EELS; slightly oxidized; zero-loss peak intensity = 5.17e9; total elastic int. / zero-loss int. = 0.21; gain change factor = 672; 70 sweeps 10 ms dwell; saturated zero-loss

 

Applications

Unlock the power of electron energy loss spectroscopy—explore cutting-edge applications, dive into expert-led webinars, and discover insights through detailed application notes that fuel scientific breakthroughs.

Looking for more Information

Want to enhance your EELS results?

Visit EM Academy to stream lessons from application specialists and design engineers anytime, anywhere, and at your own pace.